Transcatheter Repair of Anterior Mitral Leaflet Perforation Deploy, Retrieve, Redeploy

We describe a case of anterior mitral leaflet perforation successfully treated with the Amplatzer Cribriform device, with resultant hemolytic anemia. The device was retrieved, and perforation occluded with the GORE CARDIOFORM device with resolution of hemolysis. ( Level of Difficulty: Advanced. ) We...

Full description

Saved in:
Bibliographic Details
Published inJACC. Case reports Vol. 1; no. 5; pp. 689 - 693
Main Authors Panaich, Sidakpal S., Qazi, Abdul H., Horwitz, Phillip A., Staffey, Kimberly, Rossen, James D.
Format Journal Article
LanguageEnglish
Published Elsevier 18.12.2019
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:We describe a case of anterior mitral leaflet perforation successfully treated with the Amplatzer Cribriform device, with resultant hemolytic anemia. The device was retrieved, and perforation occluded with the GORE CARDIOFORM device with resolution of hemolysis. ( Level of Difficulty: Advanced. ) We describe a case of anterior mitral leaflet perforation successfully treated with the Amplatzer Cribriform device, with resultant hemolytic anemia…
ISSN:2666-0849
DOI:10.1016/j.jaccas.2019.10.014