Dose perturbation effects by metal hip prosthesis in gynaecological 192 Ir HDR brachytherapy

The present study was conducted to elucidate the effects of hip prostheses in Ir HDR brachytherapy and determine dose uncertainties introduced by the treatment planning. A gynaecological phantom irradiated using Nucletron Ir microSelectron HDR source was modeled using MCNP5 code. Three hip materials...

Full description

Saved in:
Bibliographic Details
Published inApplied radiation and isotopes Vol. 196; p. 110751
Main Authors Ab Shukor, N S, Abdullah, R, Abdul Aziz, M Z, Samson, D O, Musarudin, M
Format Journal Article
LanguageEnglish
Published England 01.06.2023
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:The present study was conducted to elucidate the effects of hip prostheses in Ir HDR brachytherapy and determine dose uncertainties introduced by the treatment planning. A gynaecological phantom irradiated using Nucletron Ir microSelectron HDR source was modeled using MCNP5 code. Three hip materials considered in this study were water, bone, and metal prosthesis. According to the obtained results, a dose perturbation was observed within the medium with a higher atomic number, which reduced the dose to the nearby region.
ISSN:1872-9800