Full-field and automatic methodology of spectral calibration for PGP imaging spectrometer
In order to analyze spectral data quantitatively which is obtained by prism-grating-prism imaging spectrometer, spectral calibration is required in order to determine spectral characteristics of PGP imaging spectrometer, such as the center wavelength of every spectral channel, spectral resolution an...
Saved in:
Published in | Guang pu xue yu guang pu fen xi Vol. 34; no. 8; p. 2289 |
---|---|
Main Authors | , , , , , |
Format | Journal Article |
Language | Chinese |
Published |
China
01.08.2014
|
Online Access | Get more information |
Cover
Loading…
Summary: | In order to analyze spectral data quantitatively which is obtained by prism-grating-prism imaging spectrometer, spectral calibration is required in order to determine spectral characteristics of PGP imaging spectrometer, such as the center wavelength of every spectral channel, spectral resolution and spectral bending. A spectral calibration system of full field based on collimated monochromatic light method is designed. Spherical mirror is used to provide collimated light, and a freely sliding and rotating folding mirror is adopted to change the angle of incident light in order to realize full field and automatic calibration of imaging spectrometer. Experiments of spectral calibration have been done for PGP imaging spectrometer to obtain parameters of spectral performance, and accuracy analysis combined with the structural features of the entire spectral calibration system have been done. Analysis results indicate that spectral calibration accuracy of the calibration system reaches 0.1 nm, and the bandwidth a |
---|---|