First-in-man successful use of the SPECTRUM percutaneous dual-stage right ventricle and right atrium to pulmonary artery ventricular assist device

BACKGROUNDOver the past decade, several minimally invasive mechanical support devices have been introduced into clinical practice to support the right ventricle (RV). Percutaneous cannulas are easy to insert, minimally invasive, and treat acute RV failure rapidly. In December 2021, the Food and Drug...

Full description

Saved in:
Bibliographic Details
Published inJournal of cardiac surgery Vol. 37; no. 10; pp. 3403 - 3407
Main Authors Usman, Asad A, Spelde, Audrey E, Olia, Salim E, Cevasco, Marisa, Bermudez, Christian, Haddle, John, Ibrahim, Michael, Szeto, Wilson, Vernick, William, Gutsche, Jacob
Format Report
LanguageEnglish
Published 01.10.2022
Online AccessGet full text

Cover

Loading…
More Information
Summary:BACKGROUNDOver the past decade, several minimally invasive mechanical support devices have been introduced into clinical practice to support the right ventricle (RV). Percutaneous cannulas are easy to insert, minimally invasive, and treat acute RV failure rapidly. In December 2021, the Food and Drug Administration approved a new 31 French dual lumen single cannula for use as a right ventricular assist device. AIMSDescirbe the use of the new dual lumen percutaneous right ventricular assist device (RVAD) cannula. MATERIAL AND METHODSDeployment of the RVAD can be done surgically or percutaneously. This cannula, manufactured by Spectrum, is dual staged. It has inflow ports positioned both in the right atrium (RA) as well as the RV for maximal drainage of the right heart. The distal end of the cannula which includes the outflow port is positioned in the pulmonary artery (PA). RESULTSDeployment of the Spectrum RVAD can be done percutaneously with transesophageal and flouroscopy guidence. Cannulation requires requisite wire skills in order to navigate into the main pulmonary artery. Utilization of this cannula can be done in acute RV failure secondary to ischemia, post cardiotomy shock, acute respiratory failure or other causes of isolated RV failure. DISCUSSIONThe dual stage drainage design optimizes venous drainage as well as limits suck-down events. Theoretically, direct RV decompression also decreases RV dilation and wall tension, and facilitates improved transmural pressure gradient to reduce RV strain. CONCLUSIONHere we describe the first-in-man successful use of the dual-stage RA and RV to PA Spectrum cannula in a patient with severe COVID acute respiratory distress syndrome and acute right ventricular failure, bridged to recovery.
Bibliography:ObjectType-Case Study-2
content type line 59
SourceType-Reports-1
ObjectType-Report-1
ISSN:1540-8191