Estimation of the Accuracy of Method for Quantitative Determination of Volatile Compounds in Alcohol Products

Results of the estimation of the precision for determination volatile compounds in alcohol-containing products by gas chromatography: acetaldehyde, methyl acetate, ethyl acetate, methanol, isopropyl alcohol, propyl alcohol, isobutyl alcohol, butyl alcohol, isoamyl alcohol are presented. To determine...

Full description

Saved in:
Bibliographic Details
Published inarXiv.org
Main Authors Charapitsa, S V, Zayats, N I, Zadreyko, Y V, Sytova, S N
Format Paper
LanguageEnglish
Published Ithaca Cornell University Library, arXiv.org 20.01.2016
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Results of the estimation of the precision for determination volatile compounds in alcohol-containing products by gas chromatography: acetaldehyde, methyl acetate, ethyl acetate, methanol, isopropyl alcohol, propyl alcohol, isobutyl alcohol, butyl alcohol, isoamyl alcohol are presented. To determine the accuracy, measurements were planned in accordance with ISO 5725 and held at the gas chromatograph Chromatec-Crystal 5000. Standard deviation of repeatability, intermediate precision and their limits are derived from obtained experimental data. The uncertainty of the measurements was calculated on the base of an "empirical" method. The obtained values of accuracy indicate that the developed method allows measurement uncertainty extended from 2 to 20% depending on the analyzed compound and measured concentration.
ISSN:2331-8422