Comparison of D^sub w^ measurements by alanine and synthetic diamond dosimeters in photon beams with 1 cm × 1 cm field size
For photon beam sizes lower than 2 cm x 2 cm no absorbed-dose-to-water (D...) primary standards are available for determining the possible influence of field size on the dosimeter response. In this work, D... measurements performed in 1 cm x 1 cm photon beams at PTB using alanine and synthetic diamo...
Saved in:
Published in | Metrologia Vol. 49; no. 5; p. S211 |
---|---|
Main Authors | , , , |
Format | Journal Article |
Language | English |
Published |
Bristol
IOP Publishing
01.10.2012
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | For photon beam sizes lower than 2 cm x 2 cm no absorbed-dose-to-water (D...) primary standards are available for determining the possible influence of field size on the dosimeter response. In this work, D... measurements performed in 1 cm x 1 cm photon beams at PTB using alanine and synthetic diamond dosimeters were compared in order to assess the reliability of the two dosimetric systems in such a small field size. The ratio of D... values measured by diamond dosimeters to D... values measured by alanine was 0.991 ± 1.5% (1 SD) and 1.007 ± 1.3% (1 SD) for 6 MV and 10 MV photon beams, respectively. This shows that both systems are capable of accurately measuring D... in small fields. (ProQuest: ... denotes formulae/symbols omitted.) |
---|---|
ISSN: | 0026-1394 1681-7575 |