Patient QA System Using Delta$^4$ Phantom for Tomotherapy: A Comparative Study with EBT3 Film
In this study for delivery quality assurance (DQA) of helical tomotherapy, we compared the Delta$^4$ phantom system with the Gafchromic EBT3 film dosimetry system in eight cases and analyzed the global gamma-index passing rate (GPR) of various treatment sites in 109 cases. The mean GPR values for De...
Saved in:
Published in | Journal of the Korean Physical Society pp. 816 - 821 |
---|---|
Main Authors | , , , , , , , |
Format | Journal Article |
Language | English |
Published |
한국물리학회
01.04.2019
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | In this study for delivery quality assurance (DQA) of helical tomotherapy, we compared the Delta$^4$ phantom system with the Gafchromic EBT3 film dosimetry system in eight cases and analyzed the global gamma-index passing rate (GPR) of various treatment sites in 109 cases. The mean GPR values for Delta$^4$ and EBT3 film were 98.8\% and 97.8\%, respectively. GPR values were similar between Delta$^4$ and film dosimetry, with a linear correlation factor of 72.4\%. About 70\% of the Delta$^4$ cases had GPR values greater than 99\%, although the GPR value was highly dependent on the clinical target size; GPR decreased approximately 3.7\% when the target size increased by 3 cm. The GPR results show that Delta$^4$ could be a better dosimetry tool than EBT3 film for tomotherapy DQA. KCI Citation Count: 0 |
---|---|
ISSN: | 0374-4884 1976-8524 |
DOI: | 10.3938/jkps.74.816 |