Development of magnetic field measurement system for AMS cyclotron

A high-accuracy magnetic field measurement device based on a cyclotron is being developed for accelerator mass spectrometry (AMS). In this study, a magnetic field measurement device consisting of a Hall probe sensor, piezo-motor, and step motor was developed to measure the magnetic field of the AMS...

Full description

Saved in:
Bibliographic Details
Published inNuclear engineering and technology Vol. 55; no. 8; pp. 3114 - 3120
Main Authors Ho Namgoong, Hyojeong Choi, Mitra Ghergherehchi, Donghyup Ha, Mustafa Mumyapan, Jong-Seo Chai, Jongchul Lee, Hoseung Song
Format Journal Article
LanguageKorean
Published 2023
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:A high-accuracy magnetic field measurement device based on a cyclotron is being developed for accelerator mass spectrometry (AMS). In this study, a magnetic field measurement device consisting of a Hall probe sensor, piezo-motor, and step motor was developed to measure the magnetic field of the AMS cyclotron magnet. The Hall probe sensor was calibrated to achieve positional accuracy by using polar coordinates. The measurement results between the ratchet gear and piezo-motor, which are the instruments used for driving the measurement device, were analyzed. The measurement result of the device with a piezo-motor exhibits a difference of 5 Gauss (0.04%) as compared with the simulation result.
Bibliography:KISTI1.1003/JNL.JAKO202324772094062
ISSN:1738-5733
2234-358X