GC에서 검출한계 결정을 위한 새로운 접근 방법에 대한 연구

The purity methods to determine LOD/LOQ using standard deviation of the residual, intercept and blank by IUPAC and ACS describe many of the pitfalls and pose significant challenges to analytical chemists. Therefore, the aim of this study is the development of the simple, easy, convenient and statist...

Full description

Saved in:
Bibliographic Details
Published in한국산업위생학회지 Vol. 20; no. 4; pp. 217 - 224
Main Authors 오도석, 신경애, 이지아, 임종호, 신미선, Oh, Doe-Suk, Shin, Kyoung-Ae, Lee, Ji-A, Lym, Jong-Ho, Shin, Mi-Sun
Format Journal Article
LanguageKorean
Published 2010
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:The purity methods to determine LOD/LOQ using standard deviation of the residual, intercept and blank by IUPAC and ACS describe many of the pitfalls and pose significant challenges to analytical chemists. Therefore, the aim of this study is the development of the simple, easy, convenient and statistically significant method to determine LOD in quantitative analysis of organic solvents by GC. The new approaching method by linearization in the given concentration range used coefficient of variation ; ${\sigma}_{n-1}$/S(standard deviation, ${\sigma}_{n-1}$ and average, S) of sensitivity(Response/concentration). The comparison of results among the purity methods(IUPAC and ACS) and the linearization have been fulfilled the F-test for standard deviations and t-test for LOD range values. The results of F-test and t-test are satisfied within 95 % confidence level, respectably. The LOD values determined by the new procedure are n-Hexane 0.0116 mg/$m^3$, Toluene 0.0807 mg/$m^3$, and o-Xylene 0.0494 mg/$m^3$. Because the standard deviation of the residual, intercept and blank and the slope of calibration curve are not calculated and the new approaching method use the coefficient of variation of sensitivity by linearization, this new method is simple, easy, convenient and statistically significant. In future, many chemical analysts will expect to applicate and routinely use this method in the all quantitative analysis.
Bibliography:KISTI1.1003/JNL.JAKO201035643038945
ISSN:1226-4326