A New PVC-Membrane Electrode Based on a Thia-Substituted Macrocyclic Diamide for Selective Potentiometric Determination of Silver Ion

A new PVC-membrane electrode for $Ag^+$ ion based on a thia-substituted macrocyclic diamide has been prepared. The electrode exhibited a Nernstian response for $Ag^+$ over a wide concentration range $(1.7{\times}10^{-6}-1.0{\times}10^{-1}M)$. It has a response time <15 s and can be used for at le...

Full description

Saved in:
Bibliographic Details
Published inBulletin of the Korean Chemical Society Vol. 23; no. 1; pp. 53 - 58
Main Authors Shamsipur, Mojtaba, Kazemi, Sayed Yahya, Niknam, Khodabaksh, Sharghi, Hashem
Format Journal Article
LanguageKorean
Published 2002
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:A new PVC-membrane electrode for $Ag^+$ ion based on a thia-substituted macrocyclic diamide has been prepared. The electrode exhibited a Nernstian response for $Ag^+$ over a wide concentration range $(1.7{\times}10^{-6}-1.0{\times}10^{-1}M)$. It has a response time <15 s and can be used for at least 3 months without divergence. The proposed membrane sensor revealed good selectivities for $Ag^+$ over a variety of metal ions and can be used in a pH range 3.0-7.5. It has been used successfully for direct determination of $Ag^+$ in different real samples and, as an indicator electrode, in the titration of silver ion.
Bibliography:KISTI1.1003/JNL.JAKO200202727371049
ISSN:0253-2964
1229-5949