A sampling multichannel bioimpedance analyzer for tissue monitoring
The paper focuses on principles of designing of a multichannel bioimpedance analyzer based on simultaneous multisine measurement. The measurement task arises due to the need to monitor patients during and after heart surgery operation performing MIMO (multiple-input-multiple-output) bioimpedance mea...
Saved in:
Published in | The 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society Vol. 1; pp. 902 - 905 |
---|---|
Main Authors | , , , , |
Format | Conference Proceeding |
Language | English |
Published |
IEEE
2004
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | The paper focuses on principles of designing of a multichannel bioimpedance analyzer based on simultaneous multisine measurement. The measurement task arises due to the need to monitor patients during and after heart surgery operation performing MIMO (multiple-input-multiple-output) bioimpedance measurement. Frequencies of the simultaneously applied sinusoidal excitations must be close but simultaneously varied in a larger range (e.g. from 1 kHz up to 10 MHz). The main idea of the proposed approach is that the use of a rather specific signal system (frequencies of sinusoidal excitations are related as integers and sampling frequencies are properly related/adapted to them) makes it possible to separate responses to different excitations from the measured summary signals by means of a quite simple filter and different (under) sampling rates. |
---|---|
ISBN: | 0780384393 9780780384392 |
DOI: | 10.1109/IEMBS.2004.1403305 |