Sensitive amperometric determination of chemical oxygen demand using Ti/Sb–SnO 2/PbO 2 composite electrode
A novel Ti/Sb–SnO 2/PbO 2 composite electrode was fabricated for COD determination. The new electrode configuration improved the sensitivity of the amperometric method apparently. Effects of common experimental parameters, such as applied potential, pH and concentration of the electrolyte on its ana...
Saved in:
Published in | Sensors and actuators. B, Chemical Vol. 155; no. 1; pp. 114 - 119 |
---|---|
Main Authors | , , , , |
Format | Journal Article |
Language | English |
Published |
Elsevier B.V
2011
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | A novel Ti/Sb–SnO
2/PbO
2 composite electrode was fabricated for COD determination. The new electrode configuration improved the sensitivity of the amperometric method apparently. Effects of common experimental parameters, such as applied potential, pH and concentration of the electrolyte on its analytical performance were investigated. A linear range of 0.5–200
mg
L
−1 COD and a detection limit (a signal-to-noise ratio of 3) of 0.3
mg
L
−1 were achieved under optimized conditions. The experiments for detecting COD in model samples and real samples were carried out to evaluate the electrode's performance. The obtained results were in good agreement with those determined by the standard dichromate method, with a relative error less than 12%. |
---|---|
ISSN: | 0925-4005 1873-3077 |