Performance evaluation of an API Aerosizer TM

A numerical method was developed to simulate the compressible flow field and particle trajectory in an API Aerosizer TM. Experimental particle time-of-flights (τ TOF) data using monodisperse solid and liquid particles were also obtained. Theoretical particle τ TOF results were compared with the pres...

Full description

Saved in:
Bibliographic Details
Published inJournal of aerosol science Vol. 29; no. 7; pp. 839 - 853
Main Authors Tsai, Chuen Jinn, Chein, Hung Min, Chang, Shu Ting, Kuo, Jong Yoh
Format Journal Article
LanguageEnglish
Published Elsevier Ltd 1998
Online AccessGet full text

Cover

Loading…
More Information
Summary:A numerical method was developed to simulate the compressible flow field and particle trajectory in an API Aerosizer TM. Experimental particle time-of-flights (τ TOF) data using monodisperse solid and liquid particles were also obtained. Theoretical particle τ TOF results were compared with the present experimental data and those of Cheng et al. (1993, J. Aerosol Sci. 26, 501–514). Good agreement was found for solid particles. The original τ TOF calibration for the API Aerosizer TM was found to underestimate measured τ TOF mainly because of the assumption of one-dimensional flow in the nozzle, and incorrect use of the particle drag coefficient in the supersonic flow field. Effects of decreasing ambient pressure and increasing particle density on increasing τ TOF were found to be important.
ISSN:0021-8502
1879-1964
DOI:10.1016/S0021-8502(97)00433-3