Durability of a bovine pericardial aortic bioprosthesis based on Valve Academic Research Consortium-3 echocardiographic criteriaCentral MessagePerspective
Objectives: The Carpentier-Edwards Perimount Magna Ease (Edwards Lifesciences) pericardial bioprosthesis has demonstrated satisfying hemodynamics at midterm follow-up, but its durability remains unclear. We report our 10-year experience with this third-generation valve implanted in the aortic positi...
Saved in:
Published in | JTCVS open Vol. 11; pp. 72 - 80 |
---|---|
Main Authors | , , , , , , |
Format | Journal Article |
Language | English |
Published |
Elsevier
01.09.2022
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Objectives: The Carpentier-Edwards Perimount Magna Ease (Edwards Lifesciences) pericardial bioprosthesis has demonstrated satisfying hemodynamics at midterm follow-up, but its durability remains unclear. We report our 10-year experience with this third-generation valve implanted in the aortic position, with particular attention to structural valve deterioration. Methods: From 2007 to 2016 at our center, 338 patients underwent aortic valve replacement using the Perimount Magna Ease pericardial bioprosthesis. Patients were prospectively followed (mean 6.6 ± 2.6 years) with clinical evaluation and yearly echocardiography. Follow-up was 98% complete (7 patients lost) for a total of 2238 valve-years. Bioprosthesis structural valve deterioration was determined by strict echocardiographic assessment based on the Valve Academic Research Consortium 3 criteria. Results: Overall operative mortality was 1.2%. Actuarial survival including early deaths averaged 80.9% ± 2.2% and 66.7% ± 4.4% after 5 and 10 years of follow-up, respectively. Actuarial freedom from explantation due to structural valve deterioration at 5 and 10 years was 99.6% ± 0.4% and 88.8% ± 5.0%, respectively, and actuarial freedom of structural valve deterioration at 5 and 10 years was 98.5% ± 0.7% and 44.0% ± 6.4%, respectively. More precisely, actuarial freedom of structural valve deterioration stage 3 was 99.6% ± 0.4% at 5 years and 88.3% ± 5.0% at 10 years, whereas freedom of structural valve deterioration stage 2/3 was 98.5% ± 0.7% and 60.9% ± 7.0%, respectively. Conclusions: With a low rate of explantation due to structural valve deterioration events at 10 years, and particularly a low rate of moderate or severe structural valve deterioration based on echocardiographic Valve Academic Research Consortium 3 criteria, the Carpentier-Edwards Perimount Magna Ease pericardial bioprosthesis remains a reliable choice for a tissue valve in the aortic position. Video Abstract: |
---|---|
ISSN: | 2666-2736 2666-2736 |