A 4π Multi-wire Proportional Chamber for the 4πβ-γ Coincidence Measurement
The β detector of 4πβ-γ coincidence measurement is usually a proportional counter with a single wire of which the electric field distribution is divergent and the detection area is limited, so it can only measure the activity of the point source. A 4π multi-wire proportional chamber used as the β de...
Saved in:
Published in | Tong wei su Vol. 36; no. 2; pp. 212 - 218 |
---|---|
Main Author | |
Format | Journal Article |
Language | Chinese |
Published |
Editorial Board of Journal of Isotopes
01.02.2023
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | The β detector of 4πβ-γ coincidence measurement is usually a proportional counter with a single wire of which the electric field distribution is divergent and the detection area is limited, so it can only measure the activity of the point source. A 4π multi-wire proportional chamber used as the β detector of 4πβ-γ coincidence measurement has been designed and fabricated because of the uniform electric field distribution and large detection area. The β plateau curves and the background have been tested. The results show that the plateau length of the β plateau curve is about 270 V, and the plateau slope of the upper and lower anode wires are 8.3‰ and 9.6‰ (60Co), 8.3‰ and 8.9‰ (90Sr-90Y), respectively. The background counting rate is within 10 s-1. The 4π multiwire proportional chamber can be used as the β detector of 4πβ-γ coincidence measurement for further research. |
---|---|
ISSN: | 1000-7512 |
DOI: | 10.7538/tws.2022.youxian.054 |