A High Precision Accelerometer Calibration Method

Centrifuge calibration can continuously provide centripetal acceleration of more than 1g, and can fully stimulate the high-order nonlinear term of accelerometer. It has become the most commonly used method to calibrate high-precision accelerometer. However, affected by the error of test equipment, t...

Full description

Saved in:
Bibliographic Details
Published inHangkong Bingqi Vol. 29; no. 5; pp. 88 - 93
Main Author Lu Kewen, Wang Xinlong, Wang Bin, Nie Guanghao, Hu Xiaodong
Format Journal Article
LanguageChinese
Published Editorial Office of Aero Weaponry 01.10.2022
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Centrifuge calibration can continuously provide centripetal acceleration of more than 1g, and can fully stimulate the high-order nonlinear term of accelerometer. It has become the most commonly used method to calibrate high-precision accelerometer. However, affected by the error of test equipment, the accuracy of the traditional centrifuge calibration model is low, which leads to the low calibration accuracy of the traditional calibration method. Therefore, a high-precision accelerometer centrifuge calibration method is proposed in this paper. By introducing the idea of semi-parametric regression in mathematical statistics, the system error caused by the error of test equipment is regarded as a non-parametric vector, and a high-precision semi-parametric calibration model including the error of test equipment is established. Then the natural spline semi-parametric estimation method is used to estimate the calibration parameters and the non-parametric vector, so as to eliminate the influence of test equipment e
ISSN:1673-5048
DOI:10.12132/ISSN.1673-5048.2022.0047