Tenth value layers for 60Co gamma rays and for 4, 6, 10, 15, and 18 MV x rays in concrete for beams of cone angles between 0 degrees and 14 degrees calculated by Monte Carlo simulation

The calculation of shielding barrier thicknesses for radiation therapy facilities according to the NCRP formalism is based on the use of broad beams (that is, the maximum possible field sizes). However, in practice, treatment fields used in radiation therapy are, on average, less than half the maxim...

Full description

Saved in:
Bibliographic Details
Published inHealth physics (1958) Vol. 92; no. 5; p. 456
Main Authors Jaradat, Adnan K, Biggs, Peter J
Format Journal Article
LanguageEnglish
Published United States 01.05.2007
Subjects
Online AccessGet more information

Cover

Loading…
More Information
Summary:The calculation of shielding barrier thicknesses for radiation therapy facilities according to the NCRP formalism is based on the use of broad beams (that is, the maximum possible field sizes). However, in practice, treatment fields used in radiation therapy are, on average, less than half the maximum size. Indeed, many contemporary treatment techniques call for reduced field sizes to reduce co-morbidity and the risk of second cancers. Therefore, published tenth value layers (TVLs) for shielding materials do not apply to these very small fields. There is, hence, a need to determine the TVLs for various beam modalities as a function of field size. The attenuation of (60)Co gamma rays and photons of 4, 6, 10, 15, and 18 MV bremsstrahlung x ray beams by concrete has been studied using the Monte Carlo technique (MCNP version 4C2) for beams of half-opening angles of 0 degrees , 3 degrees , 6 degrees , 9 degrees , 12 degrees , and 14 degrees . The distance between the x-ray source and the distal surface of the shielding wall was fixed at 600 cm, a distance that is typical for modern radiation therapy rooms. The maximum concrete thickness varied between 76.5 cm and 151.5 cm for (60)Co and 18 MV x rays, respectively. Detectors were placed at 630 cm, 700 cm, and 800 cm from the source. TVLs have been determined down to the third TVL. Energy spectra for 4, 6, 10, 15, and 18 MV x rays for 10 x 10 cm(2) and 40 x 40 cm(2) field sizes were used to generate depth dose curves in water that were compared with experimentally measured values.
ISSN:0017-9078
1538-5159