Computerized control with radiodiagnostic equipment in bronchopulmonary investigations

The authors propose a procedure and equipment for control by using a rotation table of a PYM-20M X-ray system with an image intensifier. A 4 x 4-cm beam field is used. The procedure makes it possible to lower X-ray loads in patients by 10-15 times, to calculate the absorbed integrated radiation dose...

Full description

Saved in:
Bibliographic Details
Published inMeditsinskaia tekhnika no. 6; p. 3
Main Authors Konovalov, V K, Kagalovskiĭ, G M, Smetanin, A G, Leonov, S L, Kolotilin, D D, Ananév, G I, Sarapkin, V F, Agababian, P A
Format Journal Article
LanguageRussian
Published Russia (Federation) 01.11.1993
Subjects
Online AccessGet more information

Cover

Loading…
More Information
Summary:The authors propose a procedure and equipment for control by using a rotation table of a PYM-20M X-ray system with an image intensifier. A 4 x 4-cm beam field is used. The procedure makes it possible to lower X-ray loads in patients by 10-15 times, to calculate the absorbed integrated radiation dose, to enhance the quality of plane films, to draw up a protocol of studies. This increases the accuracy of X-ray anatomic instrument position. The apparatus can be remotely controlled.