Image quality in film and xero-mammography. I. Rendering of low contrast values (author's transl)

The rendering of low contrast values by various mammographic imaging systems was investigated by a semi-objective procedure. Test exposures of a line raster with variable line widths were evaluated visually. Significantly better results were achieved by a xero-radiographic method. This was better th...

Full description

Saved in:
Bibliographic Details
Published inRöFo : Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebende Verfahren Vol. 126; no. 6; p. 529
Main Authors Säbel, M, Paterok, E M, Weishaar, J, Willgeroth, F
Format Journal Article
LanguageGerman
Published Germany 01.06.1977
Subjects
Online AccessGet more information

Cover

Loading…
More Information
Summary:The rendering of low contrast values by various mammographic imaging systems was investigated by a semi-objective procedure. Test exposures of a line raster with variable line widths were evaluated visually. Significantly better results were achieved by a xero-radiographic method. This was better than non-screen mammographic film in terms of object size, lower sensitivity to scatter and better resolution. The use of a tungsten rotating anode, with about 0.5 mm. Al filter, in xeromammography provided a satisfactory compromise between image quality and radiation dose in the rendering of low contrast studies.
ISSN:1438-9029