Measuring tissue sodium concentration: Cross‐vendor repeatability and reproducibility of 23Na‐MRI across two sites

Background Sodium MRI (23Na‐MRI)‐derived biomarkers such as total sodium concentration (TSC) have the potential to provide information on tumor cellularity and the changes in tumor microstructure that occur following therapy. Purpose To evaluate the repeatability and reproducibility of TSC measureme...

Full description

Saved in:
Bibliographic Details
Published inJournal of magnetic resonance imaging Vol. 50; no. 4; pp. 1278 - 1284
Main Authors Riemer, Frank, McHugh, Damien, Zaccagna, Fulvio, Lewis, Daniel, McLean, Mary A., Graves, Martin J., Gilbert, Fiona J., Parker, Geoff J.M., Gallagher, Ferdia A.
Format Journal Article
LanguageEnglish
Published Hoboken, USA John Wiley & Sons, Inc 01.10.2019
Wiley Subscription Services, Inc
Subjects
Online AccessGet full text
ISSN1053-1807
1522-2586
DOI10.1002/jmri.26705

Cover

More Information
Summary:Background Sodium MRI (23Na‐MRI)‐derived biomarkers such as total sodium concentration (TSC) have the potential to provide information on tumor cellularity and the changes in tumor microstructure that occur following therapy. Purpose To evaluate the repeatability and reproducibility of TSC measurements in the brains of healthy volunteers, providing evidence for the technical validation of 23Na‐MRI‐derived biomarkers. Study Type Prospective multicenter study. Subjects Eleven volunteers (32 ± 6 years; eight males, three females) were scanned twice at each of two sites. Field Strength/Sequence Comparable 3D‐cones 23Na‐MRI ultrashort echo time acquisitions at 3T. Assessment TSC values, quantified from calibration phantoms placed in the field of view, were obtained from white matter (WM), gray matter (GM), and cerebrospinal fluid (CSF), based on automated segmentation of coregistered 1H T1‐weighted images and hand‐drawn regions of interest by two readers. Statistical Tests Coefficients of variation (CoVs) from mean TSC values were used to assess intrasite repeatability and intersite reproducibility. Results Mean GM TSC concentrations (52.1 ± 7.1 mM) were ∼20% higher than for WM (41.8 ± 6.7 mM). Measurements were highly repeatable at both sites with mean scan–rescan CoVs between volunteers and regions of 2% and 4%, respectively. Mean intersite reproducibility CoVs were 3%, 3%, and 6% for WM, GM, and CSF, respectively. Data Conclusion These results demonstrate technical validation of sodium MRI‐derived biomarkers in healthy volunteers. We also show that comparable 23Na imaging of the brain can be implemented across different sites and scanners with excellent repeatability and reproducibility. Level of Evidence: 1 Technical Efficacy: Stage 2 J. Magn. Reson. Imaging 2019;50:1278–1284.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 14
The first two and last two authors contributed equally to this work.
Contract grant sponsor: Cancer Research UK; Contract grant numbers: C8742/A18097 and C19212/A16628. This is a contribution from the Cancer Imaging Centre in Cambridge & Manchester, which is funded by the Engineering and Physical Sciences Research Council and Cancer Research UK. The project was supported by the Addenbrooke's Charitable Trust, the National Institute for Health Research (NIHR) [Cambridge Biomedical Research Centre at the Cambridge University Hospitals NHS Foundation Trust] in partnership with the University of Cambridge, Evelyn Trust, the MS Society, Cambridge Experimental Cancer Medicine Centre, and a research agreement between Philips Healthcare and the University of Manchester. The views expressed are those of the authors and not necessarily those of the NHS, the NIHR, or the Department of Health and Social Care.
ISSN:1053-1807
1522-2586
DOI:10.1002/jmri.26705