Description of the methodology for dosimetric quantification in treatments with 177Lu-DOTATATE
Implementation of dosimetry calculations in the daily practice of Nuclear Medicine Departments is, at this time, a controversial issue, partly due to the lack of a standardized methodology that is accepted by all interested parties (patients, nuclear medicine physicians and medical physicists). Howe...
Saved in:
Published in | Revista Española de medicina nuclear e imagen molecular (English ed.) Vol. 40; no. 3; pp. 167 - 178 |
---|---|
Main Authors | , , , , , , |
Format | Journal Article |
Language | English |
Published |
Spain
01.05.2021
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Implementation of dosimetry calculations in the daily practice of Nuclear Medicine Departments is, at this time, a controversial issue, partly due to the lack of a standardized methodology that is accepted by all interested parties (patients, nuclear medicine physicians and medical physicists). However, since the publication of RD 601/2019 there is a legal obligation to implement it, despite the fact that it is a complex and high resource consumption procedure. The aim of this article is to review the theoretical bases of in vivo dosimetry in treatments with 177Lu-DOTATATE. The exposed methodology is the one proposed by the MIRD Committee (Medical Internal Radiation Dose) of the SNMMI (Society of Nuclear Medicine & Molecular Imaging). According to this method, the absorbed dose is obtained as the product of 2factors: the time-integrated activity of the radiopharmaceutical present in a source region and a geometrical factor S. This approach, which a priori seems simple, in practice requires several SPECT/CT acquisitions, several measurements of the whole body activity and taking several blood samples, as well as hours of image processing and computation. The systematic implementation of these calculations, in all the patients we treat, will allow us to obtain homogeneous data to correlate the absorbed doses in the lesions with the biological effect of the treatment. The final purpose of the dosimetry calculations is to be able to maximize the therapeutic effect in the lesions, controlling the radiotoxicity in the organs at risk. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |