Analysis of saxitoxin in urine by continuous-flow fast-atom bombardment mass spectrometry

An improved method of saxitoxin analysis in urine using continuous-flow fast-atom bombardment mass spectrometry was developed. Parameters studied were matrix composition, matrix flow, temperature of probe tip, probe-tip design and sample extraction. Optimal detection was obtained using the following...

Full description

Saved in:
Bibliographic Details
Published inRapid communications in mass spectrometry Vol. 6; no. 2; p. 128
Main Authors Mirocha, C J, Cheong, W, Mirza, U, Kim, Y B
Format Journal Article
LanguageEnglish
Published England 01.02.1992
Subjects
Online AccessGet more information

Cover

Loading…
More Information
Summary:An improved method of saxitoxin analysis in urine using continuous-flow fast-atom bombardment mass spectrometry was developed. Parameters studied were matrix composition, matrix flow, temperature of probe tip, probe-tip design and sample extraction. Optimal detection was obtained using the following matrix composition: 5% glycerol, 0.5% acetic acid, 0.025% sodium dodecylsulfate, 0.1% polyethylene glycol (PEG) 400 and 0.5% PEG 300; probe-tip temperature: (approximately 55 degrees C); flow rate: 5 or 8 microL per min.; probe tip: Olson-Hogge design. The STX standard was detected at 200 pg with signal-to-noise ratio of 11. The percent recovery of saxitoxin from human urine after clean-up on a weak cation exchange column was 75%.
ISSN:0951-4198
DOI:10.1002/rcm.1290060211