Study on Impedance of Implantable Cardiac Pacemaker in Unipolar/Bipolar Pacing Mode by in Vitro Experiment

The unipolar/bipolar pacing mode of pacemaker is related to its circuit impedance, which affects the battery life. In this study, the experiment scheme of pacemaker circuit impedance test was constructed. The human blood environment was simulated by NaCl solution, and the experimental environment te...

Full description

Saved in:
Bibliographic Details
Published inZhongguo yi liao qi xie za zhi Vol. 46; no. 3; p. 237
Main Authors Ding, Ding, Lin, Kai-Bin, Huang, Dong, Guo, Xin-Wei, Wang, Yan-Peng, Li, Shuai, Li, Jing-Bo, Niu, Jin-Hai
Format Journal Article
LanguageChinese
Published China 30.05.2022
Subjects
Online AccessGet more information

Cover

Loading…
More Information
Summary:The unipolar/bipolar pacing mode of pacemaker is related to its circuit impedance, which affects the battery life. In this study, the experiment scheme of pacemaker circuit impedance test was constructed. The human blood environment was simulated by NaCl solution, and the experimental environment temperature was controlled by water bath. The results of experiments showed that under the experimental conditions similar to clinical human parameters, the difference between the circuit impedance of bipolar mode and unipolar mode is 120~200 Ω. The results of the experiment confirmed that the circuit impedance of bipolar circuit was larger than that of unipolar mode, which was found in clinical practice. The results of this study have reference value to the optimization of pacing mode and the reduction of pacemaker power consumption.
ISSN:1671-7104
DOI:10.3969/j.issn.1671-7104.2022.03.001