Research of Methods to Reduce Alarm Fatigue of Monitoring System
In order to solve alarm fatigue, the algorithm optimization strategies were researched to reduce false and worthless alarms. A four-lead arrhythmia analysis algorithm, a multiparameter fusion analysis algorithm, an intelligent threshold reminder, a refractory period delay technique were proposed and...
Saved in:
Published in | Zhongguo yi liao qi xie za zhi Vol. 44; no. 6; p. 481 |
---|---|
Main Authors | , , , , , , |
Format | Journal Article |
Language | Chinese |
Published |
China
08.12.2020
|
Subjects | |
Online Access | Get more information |
Cover
Loading…
Summary: | In order to solve alarm fatigue, the algorithm optimization strategies were researched to reduce false and worthless alarms.
A four-lead arrhythmia analysis algorithm, a multiparameter fusion analysis algorithm, an intelligent threshold reminder, a refractory period delay technique were proposed and tested with collected 28 679 alarms in multi-center study.
The sampling survey indicate that the 80.8% of arrhythmia false alarms were reduced by the four-lead analysis, the 55.9% of arrhythmia and pulse false alarms were reduced by the multi-parameter fusion analysis, the 28.0% and 29.8% of clinical worthless alarms were reduced by the intelligent threshold and refractory period delay techniques respectively. Finally, the total quantity of alarms decreased to 12 724.
To increase the dimensionality of parametric analysis and control the alarm limits and delay time are conducive to reduce alarm fatigue in intensive care units. |
---|---|
ISSN: | 1671-7104 |
DOI: | 10.3969/j.issn.1671-7104.2020.06.003 |