Research of Methods to Reduce Alarm Fatigue of Monitoring System

In order to solve alarm fatigue, the algorithm optimization strategies were researched to reduce false and worthless alarms. A four-lead arrhythmia analysis algorithm, a multiparameter fusion analysis algorithm, an intelligent threshold reminder, a refractory period delay technique were proposed and...

Full description

Saved in:
Bibliographic Details
Published inZhongguo yi liao qi xie za zhi Vol. 44; no. 6; p. 481
Main Authors Liu, Mengxing, Sun, Zehui, Ye, Wenyu, Liu, Sanchao, He, Xianliang, Wang, Cheng, Li, Ye
Format Journal Article
LanguageChinese
Published China 08.12.2020
Subjects
Online AccessGet more information

Cover

Loading…
More Information
Summary:In order to solve alarm fatigue, the algorithm optimization strategies were researched to reduce false and worthless alarms. A four-lead arrhythmia analysis algorithm, a multiparameter fusion analysis algorithm, an intelligent threshold reminder, a refractory period delay technique were proposed and tested with collected 28 679 alarms in multi-center study. The sampling survey indicate that the 80.8% of arrhythmia false alarms were reduced by the four-lead analysis, the 55.9% of arrhythmia and pulse false alarms were reduced by the multi-parameter fusion analysis, the 28.0% and 29.8% of clinical worthless alarms were reduced by the intelligent threshold and refractory period delay techniques respectively. Finally, the total quantity of alarms decreased to 12 724. To increase the dimensionality of parametric analysis and control the alarm limits and delay time are conducive to reduce alarm fatigue in intensive care units.
ISSN:1671-7104
DOI:10.3969/j.issn.1671-7104.2020.06.003