The use of the deuterated internal standard for morphine quantitation for the purpose of doping control by gas chromatography with mass-selective detection
The objective of this study was to demonstrate the possibility to use deuterated compounds as internal standards for the quantitative analysis of morphine by gas chromatography with mass-selective detection for the purpose of doping control. The paper is focused on the problems associated with the u...
Saved in:
Published in | Sudebno-meditsinskaia ekspertiza Vol. 53; no. 3; p. 29 |
---|---|
Main Authors | , , , , |
Format | Journal Article |
Language | Russian |
Published |
Russia (Federation)
01.05.2010
|
Subjects | |
Online Access | Get more information |
Cover
Loading…
Summary: | The objective of this study was to demonstrate the possibility to use deuterated compounds as internal standards for the quantitative analysis of morphine by gas chromatography with mass-selective detection for the purpose of doping control. The paper is focused on the problems associated with the use of deuterated morphine-D3 as the internal standard. Quantitative characteristics of the calibration dependence thus documented are presented along with uncertainty values obtained in the measurements with the use of deuterated morphine-D6. An approach to the assessment of method bias associated with the application of morphine-D6 as the deuterated internal standard is described. |
---|---|
ISSN: | 0039-4521 |