Biomechanical effect of the iliac screw insertion depth on lumbo-iliac fixation construct

To biomechanically compare the stability of the short and long iliac screw fixation constructs in lumbo-iliac reconstruction. Seven adult human embalmed cadavers (L(3)-pelvis) were used. Using posterior spinal fixation system, L(4)-S(1) pedicle screw fixation was performed. This was defined as intac...

Full description

Saved in:
Bibliographic Details
Published inChung-hua wai kʿo tsa chih Vol. 46; no. 15; p. 1179
Main Authors Chen, Hui, Yu, Bin-sheng, Zheng, Zhao-min, Lü, You, Zhang, Kui-bo, Liu, Hui, Li, Fo-bao
Format Journal Article
LanguageChinese
Published China 01.08.2008
Subjects
Online AccessGet more information

Cover

Loading…
More Information
Summary:To biomechanically compare the stability of the short and long iliac screw fixation constructs in lumbo-iliac reconstruction. Seven adult human embalmed cadavers (L(3)-pelvis) were used. Using posterior spinal fixation system, L(4)-S(1) pedicle screw fixation was performed. This was defined as intact state of the sacroiliac joint. After the intact test, total sacrum resection and L(4)-L(5)-pelvis reconstruction by pedicle screw and iliac screw with different lengths were performed as follow: short screw group (as the length of exceeding 2 mm over ischial notch) and long screw group (as the length of exceeding 2 mm over anterior inferior iliac spine). Using the 858 MTS material testing machine, biomechanical testing was performed under 800 N compression and 7 Nm torsion loading modes. At last, the axial pullout test of two iliac screws was executed. Construct stiffness in compression and torsion test, and maximum pullout force were analyzed. Insertion lengths of the short and long iliac screw were (70 +/- 2) m
ISSN:0529-5815