Test-bench calibration of the Double-Slit Ion Energy Analyzer for Heavy Ion Beam Probing Diagnostics
This work presents the results of the design study, investigations of the features and a calibration of double-slit ion energy analyzer prototype at the test-bench in IPP NSC KIPT, Kharkov, Ukraine, with beam energy up to 10 keV. Hitherto obtained accuracy for both analyzer slits is DeltaG-2X10-3 an...
Saved in:
Published in | AIP conference proceedings Vol. 993; no. 1; pp. 239 - 242 |
---|---|
Main Authors | , , , , , , , , , , , |
Format | Journal Article |
Language | English |
Published |
United States
01.01.2008
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | This work presents the results of the design study, investigations of the features and a calibration of double-slit ion energy analyzer prototype at the test-bench in IPP NSC KIPT, Kharkov, Ukraine, with beam energy up to 10 keV. Hitherto obtained accuracy for both analyzer slits is DeltaG-2X10-3 and DeltaF-3X10-3, which is limited by the existing step motor accuracy 0.1 mm for entrance slits moving. The presented study work is considered as the first step to the multi-slit analyzer elaboration. |
---|---|
Bibliography: | SourceType-Scholarly Journals-2 ObjectType-Feature-2 ObjectType-Conference Paper-1 content type line 23 SourceType-Conference Papers & Proceedings-1 ObjectType-Article-3 |
ISBN: | 9780735405127 0735405123 |
ISSN: | 0094-243X 1551-7616 |
DOI: | 10.1063/1.2909119 |