Detection of subependymal veins using high-resolution magnetic resonance venography
High-resolution magnetic resonance venography (HR-MRV) of intracranial subependymal veins using a two-dimensional Fourier-transform time-of-flight technique was performed on normal volunteers and clinical cases of cerebral disease. For the pulse sequence, fast-field-echo sequence was used with the f...
Saved in:
Published in | Acta medica Okayama Vol. 51; no. 6; pp. 321 - 325 |
---|---|
Main Authors | , , , , , , , |
Format | Journal Article |
Language | English |
Published |
Japan
01.12.1997
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | High-resolution magnetic resonance venography (HR-MRV) of intracranial subependymal veins using a two-dimensional Fourier-transform time-of-flight technique was performed on normal volunteers and clinical cases of cerebral disease. For the pulse sequence, fast-field-echo sequence was used with the following parameters: TR/TE/flip angle = 34 ms/12 ms/50 deg., 256 x 256 matrix, 1 mm effective slice thickness, 150 mm field of view, and one signal acquisition. Sequential vertical coronal sections were taken against the skull base. The anterior septal vein, the medial atrial vein, the anterior caudate vein and thalamostriate vein were detected in all subjects. In all clinical cases, HR-MRV was equal in diagnostic capability to conventional cerebral angiography. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 0386-300X |