집속초음파 자극기의 성능평가를 위한 팬텀 내부온도 측정

This research is to measure real-time temperature distribution inside a tissue-mimicking phantom for the safety and effectiveness evaluations of focused ultrasound (FUS) device capable of linear scanning stimulation. Since the focusing area of the FUS stimulation device is smaller than diameter of c...

Full description

Saved in:
Bibliographic Details
Published inJournal of biomedical engineering research Vol. 43; no. 3; pp. 147 - 152
Main Authors 도일, 조주형, 김성목, 백경민, 김용태, 박승민, Doh, Il, Joe, Daniel J, Kim, Sung Mok, Baik, Kyung Min, Kim, Yong Tae, Park, Seung Min
Format Journal Article
LanguageKorean
Published 대한의용생체공학회 01.06.2022
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:This research is to measure real-time temperature distribution inside a tissue-mimicking phantom for the safety and effectiveness evaluations of focused ultrasound (FUS) device capable of linear scanning stimulation. Since the focusing area of the FUS stimulation device is smaller than diameter of conventional thermal probe and keeps moving, it is impossible to monitor temperature distribution inside the phantom. By using the phantom with a thin film temperature sensor array inserted, real-time temperature change caused by the FUS device was measured. The translation of the measured temperature peak was also tracked successfully. The present phantom had been experimentally proven to be applicable to validate the performance and safety of the therapeutic ultrasound devices.
Bibliography:KISTI1.1003/JNL.JAKO202219957332255
ISSN:1229-0807
2288-9396