집속초음파 자극기의 성능평가를 위한 팬텀 내부온도 측정
This research is to measure real-time temperature distribution inside a tissue-mimicking phantom for the safety and effectiveness evaluations of focused ultrasound (FUS) device capable of linear scanning stimulation. Since the focusing area of the FUS stimulation device is smaller than diameter of c...
Saved in:
Published in | Journal of biomedical engineering research Vol. 43; no. 3; pp. 147 - 152 |
---|---|
Main Authors | , , , , , , , , , , , |
Format | Journal Article |
Language | Korean |
Published |
대한의용생체공학회
01.06.2022
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | This research is to measure real-time temperature distribution inside a tissue-mimicking phantom for the safety and effectiveness evaluations of focused ultrasound (FUS) device capable of linear scanning stimulation. Since the focusing area of the FUS stimulation device is smaller than diameter of conventional thermal probe and keeps moving, it is impossible to monitor temperature distribution inside the phantom. By using the phantom with a thin film temperature sensor array inserted, real-time temperature change caused by the FUS device was measured. The translation of the measured temperature peak was also tracked successfully. The present phantom had been experimentally proven to be applicable to validate the performance and safety of the therapeutic ultrasound devices. |
---|---|