In high energy x-ray beams, neutrons are produced due to photo - d isintegration in the target material, beam flattening filter and collimator. Neutron dose equivalents from a Mevatron were measured using two kinds of detectors. One of the detectors was a neutton dose rate meter, generally called a...
Saved in:
Published in | Japan Radiological Physics Vol. 2; no. 2; pp. 103 - 115 |
---|---|
Main Authors | , , , , , , , , , , , |
Format | Journal Article |
Language | Japanese |
Published |
Japan Society of Medical Physics
01.12.1982
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | In high energy x-ray beams, neutrons are produced due to photo - d isintegration in the target material, beam flattening filter and collimator. Neutron dose equivalents from a Mevatron were measured using two kinds of detectors. One of the detectors was a neutton dose rate meter, generally called a rem counter. The other was a track-etch detector made up of 10B and polycarbonate plates for detecting alpha-track from 10B(n, a) reaction. The track-etch detector as well as BF3 counter of remcounter is mounted inside a moderator. After the irradiation at the level of 2 to 10 mrem, polycarbonate plates are electrochemically etched to record only alpha-track as etch pits. Neutron dose equivalents were measured as a function of the distance away from the isocenter. This relation may be represented by an exponential function. The distances of every 1.3 in and 1.4 m decrease the neutron dose equivalent about one half for 10 MV and 15 MV x-ray beams respectively. It can be concluded that these two dosimetry combination provides a new approach with a number of advantages compared to some other existing techniques for such measurements of neutron dose equivalent in high energy x-ray beams. |
---|---|
ISSN: | 0288-5506 2186-9847 |
DOI: | 10.11323/jjmp1981.2.2_103 |