Effect of ethylenediaminetetraacetic acid (EDTA) on the determination of tacrolimus concentration in whole blood by the ACMIA method: Comparison between improved and conventional reagents

We previously reported that the results of improved dimension tacrolimus (TAC) assay (Siemens), which is based on the affinity column-mediated immune assay (ACMIA), are affected by EDTA used as an anticoagulant. However, EDTA was improved as a reagent to EDTA-2K in 2018. In this study, we evaluated...

Full description

Saved in:
Bibliographic Details
Published inJapanese Journal of Medical Technology Vol. 68; no. 3; pp. 514 - 518
Main Authors KANOH, Yuhsaku, KATSUMATA, Minami, MUNEKATA, Shinichi, TSUKUSHI, Tomoaki, UCHIDA, Kazuhiro, SHINOHARA, Ryota, INOUE, Kina, NAKAZAKI, Hiroko
Format Journal Article
LanguageJapanese
Published Japanese Association of Medical Technologists 25.07.2019
一般社団法人 日本臨床衛生検査技師会
Subjects
Online AccessGet full text
ISSN0915-8669
2188-5346
DOI10.14932/jamt.19-19

Cover

More Information
Summary:We previously reported that the results of improved dimension tacrolimus (TAC) assay (Siemens), which is based on the affinity column-mediated immune assay (ACMIA), are affected by EDTA used as an anticoagulant. However, EDTA was improved as a reagent to EDTA-2K in 2018. In this study, we evaluated the analytical performance of the improved TAC assay using EDTA-2K in comparison with the conventional assay. In the EDTA interference study, the TAC concentration was decreased by 16.6% with 3.8 mg/mL EDTA-2K, corresponding to a half volume in the conventional assay, and the effect was observed in a concentration-dependent manner. However, the TAC concentration was decreased by only 1.4% with 3.8 mg/mL EDTA-2K in the improved assay. The within-run precisions obtained were 4.2% at a low concentration (mean: 3.9 ng/mL), 3.7% at a medium concentration (mean: 10.9 ng/mL), and 2.8% at a high concentration (mean: 20.9 ng/mL), which were equivalent to those of the conventional assay at each concentration. In addition, the limits of quantitation of the improved and conventional assays were determined to be 0.5 and 0.3 ng/mL, respectively, confirming that the measurement accuracy of the improved assay is higher than that of the conventional assay. In conclusion, it was revealed that in the improved assay, the effect of the EDTA concentration on the blood TAC concentration measurement was reduced in comparison with that in the conventional assay. The improved assay is useful for the routine TDM of TAC.
ISSN:0915-8669
2188-5346
DOI:10.14932/jamt.19-19