心臓ペーシングシステムにおける不関電極面積の重要性
The adequacy of the surface area of indifferent plate used in analysis pacemaker system was evaluated and it was indicated that it should be more than 6.0 cm2 to obtain accurate stimulation threshold and lead impedance.In order to achieve permanent pacing with certainty, an indifferent place of enou...
Saved in:
Published in | Jinko Zoki Vol. 10; no. 4; pp. 839 - 841 |
---|---|
Main Authors | , , , , , , , , , |
Format | Journal Article |
Language | Japanese |
Published |
一般社団法人 日本人工臓器学会
1981
JAPANESE SOCIETY FOR ARTIFICIAL ORGANS |
Subjects | |
Online Access | Get full text |
ISSN | 0300-0818 1883-6097 |
DOI | 10.11392/jsao1972.10.839 |
Cover
Summary: | The adequacy of the surface area of indifferent plate used in analysis pacemaker system was evaluated and it was indicated that it should be more than 6.0 cm2 to obtain accurate stimulation threshold and lead impedance.In order to achieve permanent pacing with certainty, an indifferent place of enough size must be used though the life of battery may be shortened. |
---|---|
ISSN: | 0300-0818 1883-6097 |
DOI: | 10.11392/jsao1972.10.839 |