The adequacy of the surface area of indifferent plate used in analysis pacemaker system was evaluated and it was indicated that it should be more than 6.0 cm2 to obtain accurate stimulation threshold and lead impedance.In order to achieve permanent pacing with certainty, an indifferent place of enou...

Full description

Saved in:
Bibliographic Details
Published inJinko Zoki Vol. 10; no. 4; pp. 839 - 841
Main Authors 森本, 英夫, 岸田, 尚夫, 柿本, 祥太郎, 近藤, 敬一郎, 佐々木, 学, 武内, 敦郎, 西本, 孝, 大関, 道麿, 佐々木, 進次郎, 黒田, 克彦
Format Journal Article
LanguageJapanese
Published 一般社団法人 日本人工臓器学会 1981
JAPANESE SOCIETY FOR ARTIFICIAL ORGANS
Subjects
Online AccessGet full text
ISSN0300-0818
1883-6097
DOI10.11392/jsao1972.10.839

Cover

More Information
Summary:The adequacy of the surface area of indifferent plate used in analysis pacemaker system was evaluated and it was indicated that it should be more than 6.0 cm2 to obtain accurate stimulation threshold and lead impedance.In order to achieve permanent pacing with certainty, an indifferent place of enough size must be used though the life of battery may be shortened.