Evaluation of simple and easy method for the preparation of a standard gas of volatile organic compounds from their mixed solution
Round-robin tests were conducted in order to investigate a simple and easy method for preparing a standard gas mixture of volatile organic hazardous air pollutants from their mixed solution by 16 members of the Association for Environmental Analysis in Osaka. Seven compounds (acrylonitrile, dichloro...
Saved in:
Published in | BUNSEKI KAGAKU Vol. 48; no. 1; pp. 121 - 127 |
---|---|
Main Authors | , , , , , , , , , , , , , |
Format | Journal Article |
Language | Japanese |
Published |
The Japan Society for Analytical Chemistry
01.01.1999
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Round-robin tests were conducted in order to investigate a simple and easy method for preparing a standard gas mixture of volatile organic hazardous air pollutants from their mixed solution by 16 members of the Association for Environmental Analysis in Osaka. Seven compounds (acrylonitrile, dichloromethane, chloroform, 1, 2-dichloroethane, benzene, trichloroethylene, and tetrachloroethylene) were accurately weighed into a vial one by one. An aliquot of this liquid mixture was transferred into a glass vessel (1 or 21) using a microsyringe, and then vaporized. The resulting gas mixtures (about 300 ppm each) were evaluated as a first standard gas by gas chromatography. The obtained results prove this method to be accurate and precise for the preparation of a standard gas mixture. |
---|---|