3T-MR装置におけるSAR低減シーケンスの画像特性
Reduction of the specific absorption ratio (SAR) is essential in MR systems with high magnetic fields. Therefore, a variety of SAR-reduced sequences are available on modern 3T-MR systems, and some sequences can be applied under the same conditions as conventional sequences. The purpose of this study...
Saved in:
Published in | 日本放射線技術学会雑誌 Vol. 64; no. 12; pp. 1492 - 1496 |
---|---|
Main Authors | , , , , , |
Format | Journal Article |
Language | Japanese |
Published |
公益社団法人 日本放射線技術学会
2008
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Reduction of the specific absorption ratio (SAR) is essential in MR systems with high magnetic fields. Therefore, a variety of SAR-reduced sequences are available on modern 3T-MR systems, and some sequences can be applied under the same conditions as conventional sequences. The purpose of this study was to compare the imaging properties of a phantom acquired by SAR-reduced sequences with those of conventional sequences. Our data showed that effective relaxation time differed in each of the sequences, because the echo space and RF pulse of the SAR-reduced sequences varied from those of the conventional sequences. Further, the image contrast of the SAR-reduced sequences differed from the image contrast of the conventional ones. |
---|---|