SPECTデータ評価用ディジタルファントムを用いた異なる処理装置間のOS-EM再構成アルゴリズムの比較
Objective: In the OS-EM method, reconfigured images may be different among reconstruction image processors because the programs are highly arbitrary. The purpose of this study was to evaluate the difference in OS-EM algorithms among four different image processors using a digital phantom dedicated f...
Saved in:
Published in | 日本放射線技術学会雑誌 Vol. 64; no. 11; pp. 1361 - 1368 |
---|---|
Main Authors | , , , , , , , |
Format | Journal Article |
Language | Japanese |
Published |
公益社団法人 日本放射線技術学会
2008
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Objective: In the OS-EM method, reconfigured images may be different among reconstruction image processors because the programs are highly arbitrary. The purpose of this study was to evaluate the difference in OS-EM algorithms among four different image processors using a digital phantom dedicated for SPECT data evaluation. The image processors used were GMS-5500A/PI (Toshiba), GENIE Xeleris (GE), e.soft (Siemens), and Odyssey FX (Shimadzu). Methods: Multiple images were reconstructed with OS-EM fluctuating the number of subsets and iterations. A region of interest (ROI) was placed on each image. The average counts, contrast, root mean square uncertainty (%RMSU), and normalized mean squared error (NMSE) of each ROI were calculated and compared with those of different image processors. Results: There was no significant difference in contrast among the algorithms. However, the average counts and (%RMSU were significantly different among algorithms as the number of updates increased. In addition, the minimums of NMSE were also different among algorithms. Conclusion: In the OS-EM method, careful evaluation is necessary when using multiple image processors in research studies on the standardization of nuclear medicine imaging or in clinical applications. |
---|---|
ISSN: | 0369-4305 1881-4883 |
DOI: | 10.6009/jjrt.64.1361 |