Geometrical Detection Efficiency of Point and Disc Source
The geometrical efficiencies of a point and a disc source on the radiation detection by a detector with circular window were calculated by a polynomial expression. Those were calculated in the case of a source off the center line of the detector window, too. It was found that the geometrical efficie...
Saved in:
Published in | RADIOISOTOPES Vol. 22; no. 12; pp. 681 - 689 |
---|---|
Main Authors | , |
Format | Journal Article |
Language | Japanese |
Published |
Japan Radioisotope Association
15.12.1973
|
Online Access | Get full text |
ISSN | 0033-8303 1884-4111 |
DOI | 10.3769/radioisotopes.22.12_681 |
Cover
Summary: | The geometrical efficiencies of a point and a disc source on the radiation detection by a detector with circular window were calculated by a polynomial expression. Those were calculated in the case of a source off the center line of the detector window, too. It was found that the geometrical efficiency Gpof a point source at a point P (x, z) off the center line, in the case of z≥5A, can be approximately expressed as: Gp/Go=1.0-1.54 (x/z) 3+0.89 (x/z) 3 Go=1/2 (1.0-D/z) : geometrical efficiency of a point source on the center line. The geometrical efficiency Gsof a disc source (radius R) on the center line, in the case of z≥5A, can be approximately expressed as: Gs/Go=1.0-0.84 (R/z) 2+ (R/z) 3 |
---|---|
ISSN: | 0033-8303 1884-4111 |
DOI: | 10.3769/radioisotopes.22.12_681 |