Design of an endoluminal NOTES robotic system
Natural orifice transluminal endoscopic surgery, or NOTES, allows for exceedingly minimally invasive surgery but has high requirements for the dexterity and force capabilities of the tools. An overview of the ViaCath System is presented. This system is a first generation teleoperated robot for endol...
Saved in:
Published in | 2007 IEEE/RSJ International Conference on Intelligent Robots and Systems pp. 410 - 416 |
---|---|
Main Authors | , , , |
Format | Conference Proceeding |
Language | English |
Published |
IEEE
01.10.2007
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Natural orifice transluminal endoscopic surgery, or NOTES, allows for exceedingly minimally invasive surgery but has high requirements for the dexterity and force capabilities of the tools. An overview of the ViaCath System is presented. This system is a first generation teleoperated robot for endoluminal surgery and consists of a master console with haptic interfaces, slave drive mechanisms, and 6 degree-of-freedom, long-shafted flexible instruments that run alongside a standard gastroscope or colonoscope. The system was validated through animal studies. It was discovered that the devices were difficult to introduce into the GI tract and manipulation forces were insufficient. The design of a second generation system is outlined with improvements to the instrument articulation section and a steerable overtube. Results of basic evaluation tests performed on the tools are also presented. |
---|---|