Automated System for Calibration and Validation of Automotive Test Instrumentation

We describe the design and implementation of an automated calibration system for the thermocouple/voltage module of an automotive data-acquisition system. The automated calibration system tests and calibrates sixteen thermocouple/voltage signal-conditioning cards, with a combined total of 320 data c...

Full description

Saved in:
Bibliographic Details
Published in2006 World Automation Congress pp. 1 - 7
Main Authors Creasman, S.M., Pheanis, D.C.
Format Conference Proceeding
LanguageEnglish
Published IEEE 01.07.2006
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:We describe the design and implementation of an automated calibration system for the thermocouple/voltage module of an automotive data-acquisition system. The automated calibration system tests and calibrates sixteen thermocouple/voltage signal-conditioning cards, with a combined total of 320 data channels, over the data-acquisition system's operating range from -40degC through +85degC. The system collects initial measurements to document on-arrival performance, calibrates each channel, and then collects final measurements to verify each channel's post-calibration function, all within an eight-hour period. The system also automatically identifies channels that fail calibration and flags them for repair. The automated system has eliminated more than 96% of the previous labor cost for calibrations, allowing us to recover the entire cost of development and equipment in less than one year.