Evaluation and simulation of eight-element phased-array surface coil in 3.0T MRI
To evaluate the performance of a new eight-channel phased array surface coil, a standard sixteen volume coil was compared. Signal to noise ratio (SNR) and uniformity of the two coils was measured using two different sequences respectively. The images from a volunteer with the two coils were also eva...
Saved in:
Published in | Proceedings of 2012 IEEE-EMBS International Conference on Biomedical and Health Informatics pp. 659 - 662 |
---|---|
Main Authors | , , , |
Format | Conference Proceeding |
Language | English |
Published |
IEEE
01.01.2012
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | To evaluate the performance of a new eight-channel phased array surface coil, a standard sixteen volume coil was compared. Signal to noise ratio (SNR) and uniformity of the two coils was measured using two different sequences respectively. The images from a volunteer with the two coils were also evaluated. In addition, simulation analysis of the coil was done for further explanation of the coil performance. The results showed the SNR of eight-channel phased array coil was about three times of sixteen-channel volume coil but had a worse uniformity. The electromagnetic simulation experiment had a consistent result with the clinical imaging. |
---|---|
ISBN: | 9781457721762 1457721767 |
ISSN: | 2168-2194 2168-2208 |
DOI: | 10.1109/BHI.2012.6211669 |