Evaluation and simulation of eight-element phased-array surface coil in 3.0T MRI

To evaluate the performance of a new eight-channel phased array surface coil, a standard sixteen volume coil was compared. Signal to noise ratio (SNR) and uniformity of the two coils was measured using two different sequences respectively. The images from a volunteer with the two coils were also eva...

Full description

Saved in:
Bibliographic Details
Published inProceedings of 2012 IEEE-EMBS International Conference on Biomedical and Health Informatics pp. 659 - 662
Main Authors Wang Longchen, Zhu Gaojie, Wang Jie, Li Bin
Format Conference Proceeding
LanguageEnglish
Published IEEE 01.01.2012
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:To evaluate the performance of a new eight-channel phased array surface coil, a standard sixteen volume coil was compared. Signal to noise ratio (SNR) and uniformity of the two coils was measured using two different sequences respectively. The images from a volunteer with the two coils were also evaluated. In addition, simulation analysis of the coil was done for further explanation of the coil performance. The results showed the SNR of eight-channel phased array coil was about three times of sixteen-channel volume coil but had a worse uniformity. The electromagnetic simulation experiment had a consistent result with the clinical imaging.
ISBN:9781457721762
1457721767
ISSN:2168-2194
2168-2208
DOI:10.1109/BHI.2012.6211669