T1 and T2 estimation in complex domain: First results on clinical data

Purpose The article provides a practical method for generating T1 and T2 maps from multiple, sequentially acquired Spin Echo images. First results on real data for diagnostic purposes are presented and discussed in order to investigate the possible clinical usages of the proposed technique. The main...

Full description

Saved in:
Bibliographic Details
Published inConcepts in magnetic resonance. Part A, Bridging education and research Vol. 43; no. 5; pp. 166 - 176
Main Authors Baselice, Fabio, Caivano, Rocchina, Cammarota, Aldo, Ferraioli, Giampaolo, Pascazio, Vito
Format Journal Article
LanguageEnglish
Published Blackwell Publishing Ltd 01.09.2014
Subjects
Online AccessGet full text
ISSN1546-6086
1552-5023
DOI10.1002/cmr.a.21315

Cover

Loading…
More Information
Summary:Purpose The article provides a practical method for generating T1 and T2 maps from multiple, sequentially acquired Spin Echo images. First results on real data for diagnostic purposes are presented and discussed in order to investigate the possible clinical usages of the proposed technique. The main advantage of the method is its ability in identifying even very small variation in one or both relaxation time maps, allowing a more clear differentiation of tissues across the whole image, instead of highlighting only high vascularized regions, such as in contrast medium based images. Theory and Methods For the joint estimation of relaxation time maps, the method exploits the full complex acquired data instead of amplitude acquisitions. This allows having an estimator able to provide unbiased results even in case of tissues characterized by weak Signal to Noise Ratio. Results The effectiveness of the proposed approach is shown on different real case studies: a phantom dataset for quantitative analysis and three human brain datasets for diagnostic purposes. Conclusion The proposed method shows very interesting and promising results when applied to clinical datasets. The innovative technique is able to evaluate the biological topography and highlight lesions not otherwise easily distinguishable with standard examinations. © 2015 Wiley Periodicals, Inc. Concepts Magn Reson Part A 43A: 166–176, 2015.
Bibliography:ArticleID:CMRA21315
ark:/67375/WNG-8GSCDRDN-D
istex:A7924B79E32F2AA2BE1CF0593047EA3F357B280E
ISSN:1546-6086
1552-5023
DOI:10.1002/cmr.a.21315