Analysis of Polycyclic Aromatic Hydrocarbons (PAH) Using gas Chromatography and NEMS Detector
This paper describes the analysis of a mixture of 16 polycyclic aromatic hydrocarbons using gas chromatography (GC) and nano-electromechanical systems (NEMS). The performance of NEMS to detect PAHs is assessed and compared with that of commercial flame ionization detector (FID). The limit of detecti...
Saved in:
Published in | 2019 20th International Conference on Solid-State Sensors, Actuators and Microsystems & Eurosensors XXXIII (TRANSDUCERS & EUROSENSORS XXXIII) pp. 178 - 181 |
---|---|
Main Authors | , , , , , , |
Format | Conference Proceeding |
Language | English |
Published |
IEEE
01.06.2019
|
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | This paper describes the analysis of a mixture of 16 polycyclic aromatic hydrocarbons using gas chromatography (GC) and nano-electromechanical systems (NEMS). The performance of NEMS to detect PAHs is assessed and compared with that of commercial flame ionization detector (FID). The limit of detection of the GC-NEMS system ranges from 70 pg to 500 pg, which confirm the potential of NEMS to quantify high molecular weight molecules. The limitations of such a system to detect PAHs are discussed. |
---|---|
ISSN: | 2167-0021 |
DOI: | 10.1109/TRANSDUCERS.2019.8808382 |