Application of PILATUS II Detector Modules for High Resolution X-Ray Imaging Crystal Spectrometers on the Alcator C-Mod Tokamak

A new type of X-ray imaging crystal spectrometer for Doppler measurements of the radial profiles of the ion temperature and plasma rotation velocity in tokamak plasmas is presently being developed in a collaboration between various laboratories. The spectrometer will consist of a spherically bent cr...

Full description

Saved in:
Bibliographic Details
Published in2006 IEEE Nuclear Science Symposium Conference Record Vol. 6; pp. 3721 - 3723
Main Authors Bitter, M.L., Broennimann, C., Eikenberry, E.F., Hill, K.W., Ince-Cushman, A., Lee, S.G., Rice, J.E., Scott, S.
Format Conference Proceeding
LanguageEnglish
Published IEEE 01.10.2006
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:A new type of X-ray imaging crystal spectrometer for Doppler measurements of the radial profiles of the ion temperature and plasma rotation velocity in tokamak plasmas is presently being developed in a collaboration between various laboratories. The spectrometer will consist of a spherically bent crystal and a two-dimensional position sensitive detector; and it will record temporally and spatially resolved X-ray line spectra from highly-charged ions. The detector must satisfy challenging requirements with respect to count rate and spatial resolution. The paper presents the results from a recent test of a PILATUS II detector module on Alcator C-Mod, which demonstrate that the PILATUS II detector modules will satisfy these requirements.
ISBN:9781424405602
1424405602
ISSN:1082-3654
2577-0829
DOI:10.1109/NSSMIC.2006.353802