Making low level AC voltage measurements with DVM

The paper describes a method for the measurement of small alternating voltages, from 20 nV to 20 mV and higher, at frequencies up to 400 Hz. Particular attention is given to measurement at frequencies close but not equal to that of the mains. An application of the developed method to the characteriz...

Full description

Saved in:
Bibliographic Details
Published in2016 Conference on Precision Electromagnetic Measurements (CPEM 2016) pp. 1 - 2
Main Authors Gubler, G. B., Alekseeva, N. S.
Format Conference Proceeding
LanguageEnglish
Published IEEE 01.07.2016
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:The paper describes a method for the measurement of small alternating voltages, from 20 nV to 20 mV and higher, at frequencies up to 400 Hz. Particular attention is given to measurement at frequencies close but not equal to that of the mains. An application of the developed method to the characterization of reference current-to-voltage converters based on multistage current transformers is considered.
ISSN:2160-0171
DOI:10.1109/CPEM.2016.7540554