Size-Exclusion Chromatography for the Impurity Analysis of Adeno-Associated Virus Serotypes
Excitation: 280 nm; Emission: 350 nm Mobile phase: 10 mM NaH2P04, 10 mM Na2HP04, 200-400 mKCI, pH6.6(HCI) Flow rate: 0.6 mL/min Results and Discussion The upper analyte size limitation for SEC separations of proteins is approximately 100-200 nm (subvisible) depending on SEC particle size. Above thes...
Saved in:
Published in | LC GC North America Vol. 38; no. 6; pp. 367 - 368 |
---|---|
Main Authors | , |
Format | Magazine Article |
Language | English |
Published |
Monmouth Junction
Intellisphere, LLC
01.06.2020
MultiMedia Healthcare Inc |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | Excitation: 280 nm; Emission: 350 nm Mobile phase: 10 mM NaH2P04, 10 mM Na2HP04, 200-400 mKCI, pH6.6(HCI) Flow rate: 0.6 mL/min Results and Discussion The upper analyte size limitation for SEC separations of proteins is approximately 100-200 nm (subvisible) depending on SEC particle size. Above these limits, the analyte may be altered due to shear forces or trapped by the frits or packed bed of the column. [...]these subvisible entities are analyzed using dynamic light scattering and nanoparticle tracking analysis (NTA) methods, among others (2). AAV has a total protein and ssDNA molecular weight of 5500 kDa; however, the compact structure has a diameter of only 25 nm (4). [...]it was predicted that these high-efficiency SEC particles could provide the accessible pore volume needed. |
---|---|