Accuracy of Sphygmomanometers

One of the factors affecting the accuracy of readings of blood pressure is the equipment used. Defects or inaccuracy of aneroid sphygmomanometers may be source of error in blood pressure measurement. We inspected 100 sphygmomanometers for physical defects and assessed their accuracy against a standa...

Full description

Saved in:
Bibliographic Details
Published inTurkish journal of medical sciences Vol. 29; no. 4; pp. 487 - 491
Main Author BASAK, Okay
Format Journal Article
LanguageEnglish
Published Turkish Journal of Medical Sciences 01.04.1999
TÜBİTAK
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:One of the factors affecting the accuracy of readings of blood pressure is the equipment used. Defects or inaccuracy of aneroid sphygmomanometers may be source of error in blood pressure measurement. We inspected 100 sphygmomanometers for physical defects and assessed their accuracy against a standard mercury manometer at four different pressure points. 46 of the 100 sphygmomanometers were determined to be intolerant (deviation from the mercury manometer by greater than±3 mm Hg at two or more of the test points). There were faults in the inflation-deflation system of 34 sphygmomanometers. The most common physical defects were: defective pump bulb, defective rubber tubing and indicator needles not pointing to the "zero box". There was no regular preventive maintenance program in health units. We recommend that health care providers should use mercury sphygmomanometers and have at their disposal the equipment necessary to check the accuracy of manometers.
Bibliography:http://dergipark.ulakbim.gov.tr/tbtkmedical/article/view/5000032285
TTIP
ISSN:1303-6165
1300-0144