Inter- and intra-unit reliability of the COSMED K5: Implications for multicentric and longitudinal testing
To evaluate the intra-unit (RELINTRA) and inter-unit reliability (RELINTER) of two structurally identical units of the metabolic analyser K5 (COSMED, Rome, Italy) that allows to utilize either breath-by-breath (BBB) or dynamic mixing chamber (DMC) technology. Identical flow- and gas-signals were tra...
Saved in:
Published in | PloS one Vol. 15; no. 10; p. e0241079 |
---|---|
Main Authors | , , , , |
Format | Journal Article |
Language | English |
Published |
United States
Public Library of Science
23.10.2020
Public Library of Science (PLoS) |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | To evaluate the intra-unit (RELINTRA) and inter-unit reliability (RELINTER) of two structurally identical units of the metabolic analyser K5 (COSMED, Rome, Italy) that allows to utilize either breath-by-breath (BBB) or dynamic mixing chamber (DMC) technology.
Identical flow- and gas-signals were transmitted to both K5s that always operated simultaneously either in BBB- or DMC-mode. To assess RELINTRA and RELINTER, a metabolic simulator was applied to simulate four graded levels of respiration. RELINTRA and RELINTER were expressed as typical error (TE%) and Intraclass Correlation Coefficient (ICC). To assess also inter-unit differences via natural respiratory signals, 12 male athletes performed one incremental bike step test each in BBB- and DMC-mode. Inter-unit differences within biological testing were expressed as percentages.
In BBB, TE% of RELINTRA ranged 0.30-0.67 vs. RELINTER 0.16-1.39 and ICC ranged 0.57-1.00 vs. 0.09-1.00. In DMC, TE% of RELINTRA ranged 0.38-0.90 vs. RELINTER 0.03-0.86 and ICC ranged 0.22-1.00 vs. 0.52-1.00. Mean inter-unit differences ranged -2.30-2.20% (Cohen's ds (ds) 0.13-1.52) for BBB- and -0.55-0.61% (ds 0.00-0.65) for DMC-mode, respectively. Inter-unit differences for [Formula: see text] and RER were significant (p < 0.05) at each step.
Two structurally identical K5-units demonstrated accurate RELINTRA with TE < 2.0% and similar RELINTER during metabolic simulation. During biological testing, inter-unit differences for [Formula: see text] and RER in BBB-mode were higher than 2% with partially large ES in BBB. Hence, the K5 should be allocated personally wherever possible. Otherwise, e.g. in multicenter studies, a decrease in total reliability needs to be considered especially when the BBB-mode is applied. |
---|---|
Bibliography: | ObjectType-Article-2 SourceType-Scholarly Journals-1 ObjectType-Undefined-1 ObjectType-Feature-3 content type line 23 Competing Interests: Ulm University received partial funding from COSMED (Rome, Italy), two K5 metabolic analyzers were provided for free during the study period. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript. Further, this does not alter our adherence to PLOS ONE policies on sharing data and materials. |