One-Year Late Lumen Loss between A Polymer-Coated Paclitaxel-Eluting Stent (Eluvia) and a Polymer-Free Paclitaxel-Coated Stent (Zilver PTX) for Femoropopliteal Disease

Aim: Paclitaxel-eluting stents' (Eluvia and Zilver PTX) effectiveness has been recently reported for femoropopliteeal (FP) lesions. However, there is no evaluation of one-year late lumen loss (LLL). Therefore, we evaluated one-year LLL after implantation with Eluvia or Zilver PTX.Methods: This...

Full description

Saved in:
Bibliographic Details
Published inJournal of Atherosclerosis and Thrombosis Vol. 27; no. 2; pp. 164 - 171
Main Authors Soga, Yoshimitsu, Fujihara, Masahiko, Tomoi, Yusuke, Iida, Osamu, Ishihara, Takayuki, Kawasaki, Daizo, Ando, Kenji
Format Journal Article
LanguageEnglish
Published Japan Japan Atherosclerosis Society 01.02.2020
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Aim: Paclitaxel-eluting stents' (Eluvia and Zilver PTX) effectiveness has been recently reported for femoropopliteeal (FP) lesions. However, there is no evaluation of one-year late lumen loss (LLL). Therefore, we evaluated one-year LLL after implantation with Eluvia or Zilver PTX.Methods: This was a multicenter, prospective study. Patients who had symptomatic de novo lesions in the native FP artery were enrolled. The primary endpoint was one-year angiographic LLL, and the secondary endpoints were binary restenosis and target lesion revascularization (TLR) at one year.Results: From December 2015 to December 2016, 48 patients (Eluvia, 36 patients; Zilver PTX, 12 patients) were enrolled. No significant difference was found in baseline and lesion characteristics between both groups. One-year, LLL was significantly lower in the Eluvia group (0.60 {plus minus}0.80 mm) than in the Zilver PTX group (1.74 {plus minus}0.89 mm) (P=0.0003). Negative LLL was observed only in the Eluvia group (0% vs. 23%, p=0.096). The binary restenosis rate was significantly lower than in the Zilver PTX group (0% vs. 16.7%, P=0.012). The one-year TLR in the Eluvia group tended to be lower (0% vs. 8.3%, P=0.08). Stent thrombosis was not observed in either group.Conclusion: One-year LLL in the Eluvia group was significantly lower than that in the Zilver PTX group for FP lesions.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:1340-3478
1880-3873