Thick shells and medially wedged posts increase foot orthoses medial longitudinal arch stiffness: an experimental study

Background Foot orthoses (FOs) are commonly prescribed devices to attenuate biomechanical deficits and improve physical function in patients with musculoskeletal disorders. It is postulated that FOs provide their effects through the production of reaction forces at the foot‐FOs interface. An importa...

Full description

Saved in:
Bibliographic Details
Published inJournal of foot and ankle research Vol. 16; no. 1; pp. 11 - n/a
Main Authors Pelaez, Ana Sofia Tavera, Farahpour, Nader, Griffiths, Ian B., Moisan, Gabriel
Format Journal Article
LanguageEnglish
Published London BioMed Central 03.03.2023
BioMed Central Ltd
Wiley
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Background Foot orthoses (FOs) are commonly prescribed devices to attenuate biomechanical deficits and improve physical function in patients with musculoskeletal disorders. It is postulated that FOs provide their effects through the production of reaction forces at the foot‐FOs interface. An important parameter to provide these reaction forces is their medial arch stiffness. Preliminary results suggest that adding extrinsic additions to FOs (e.g., rearfoot posts) increases their medial arch stiffness. A better understanding of how FOs medial arch stiffness can be modulated by changing structural factors is necessary to better customise FOs for patients. The objectives of this study were to compare FOs stiffness and force required to lower the FOs medial arch in three thicknesses and two models (with and without medially wedged forefoot‐rearfoot posts). Methods Two models of FOs, 3D printed in Polynylon‐11, were used: (1) without extrinsic additions (mFO), and (2) with forefoot‐rearfoot posts and a 6o medial wedge (FO6MW). For each model, three thicknesses (2.6 mm, 3.0 mm, and 3.4 mm) were manufactured. FOs were fixed to a compression plate and vertically loaded over the medial arch at a rate of 10 mm/minute. Two‐way ANOVAs and Tukey post‐hoc tests with Bonferroni corrections were used to compare medial arch stiffness and force required to lower the arch across conditions. Results Regardless of the differing shell thicknesses, the overall stiffness was 3.4 times greater for FO6MW compared to mFO (p < 0.001). FOs with 3.4 mm and 3.0 mm thicknesses displayed 1.3‐ and 1.1‐ times greater stiffness than FOs with a thickness of 2.6 mm. FOs with a thickness of 3.4 mm also exhibited 1.1 times greater stiffness than FOs with a thickness of 3.0 mm. Overall, the force to lower the medial arch was up to 3.3 times greater for FO6MW than mFO and thicker FOs required greater force (p < 0.001). Conclusions An increased medial longitudinal arch stiffness is seen in FOs following the addition of 6o medially inclined forefoot‐rearfoot posts, and when the shell is thicker. Overall, adding forefoot‐rearfoot posts to FOs is significantly more efficient than increasing shell thickness to enhance these variables should that be the therapeutic aim.
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-Feature-2
content type line 23
ISSN:1757-1146
1757-1146
DOI:10.1186/s13047-023-00609-z