Impact of Implantation Technique and Plaque Morphology on Strut Embedment and Scaffold Expansion of Polylactide Bioresorbable Scaffold – Insights From ABSORB Japan Trial

Background:The optimal implantation technique for the bioresorbable scaffold (Absorb, Abbott Vascular) is still a matter of debate. The purpose of the present study was to evaluate the effect of implantation technique on strut embedment and scaffold expansion.Methods and Results:Strut embedment dept...

Full description

Saved in:
Bibliographic Details
Published inCirculation Journal Vol. 80; no. 11; pp. 2317 - 2326
Main Authors Sotomi, Yohei, Onuma, Yoshinobu, Dijkstra, Jouke, Eggermont, Jeroen, Liu, Shengnan, Tenekecioglu, Erhan, Zeng, Yaping, Asano, Taku, Winter, Robbert J. de, Popma, Jeffrey J., Kozuma, Ken, Tanabe, Kengo, Serruys, Patrick W., Kimura, Takeshi
Format Journal Article
LanguageEnglish
Published Japan The Japanese Circulation Society 2016
Subjects
Online AccessGet full text

Cover

Loading…
More Information
Summary:Background:The optimal implantation technique for the bioresorbable scaffold (Absorb, Abbott Vascular) is still a matter of debate. The purpose of the present study was to evaluate the effect of implantation technique on strut embedment and scaffold expansion.Methods and Results:Strut embedment depth and scaffold expansion index assessed by optical coherence tomography (OCT) (minimum scaffold area/reference vessel area) were evaluated in the ABSORB Japan trial (OCT subgroup: 87 lesions) with respect to implantation technique using either quantitative coronary angiography (QCA) or OCT. Strut embedment was assessed at the strut level (n=667), while scaffold expansion was assessed at the lesion level (n=81). The mean embedment depth was 63±59 µm. Balloon sizing and inflation pressure had no direct effect on strut embedment. Plaque morphology affected strut embedment [nonatherosclerotic (58.9±54.3 µm), fibroatheroma (73.3±59.6 µm), fibrous plaque (59.7±51.1 µm), and fibrocalcific plaque (–3.1±61.6 µm, negative value means malapposition), P <0.001]. The balloon-artery ratio positively correlated with the expansion index. This relationship was stronger when the OCT-derived reference vessel diameter (RVD) was used as a reference for balloon selection rather than the QCA-derived one [predilatation (Pearson correlation r: QCA: 0.167 vs. OCT: 0.552), postdilatation (QCA: 0.316 vs. OCT: 0.717)].Conclusions:Underlying plaque morphology influenced strut embedment, whereas implantation technique had no direct effect on it. Optimal balloon sizing based on OCT-derived RVD might be recommended. However, the safety of such a strategy should be investigated in a prospective trial. (Circ J 2016; 80: 2317–2326)
Bibliography:ObjectType-Article-1
SourceType-Scholarly Journals-1
ObjectType-News-2
ObjectType-Feature-3
content type line 23
ISSN:1346-9843
1347-4820