Interlaboratory Testing for the Determination of Trace Amounts of Tin and Lead in Magnesium and Magnesium Alloys by Inductively Coupled Plasma Atomic Emission Spectrometry
In order to establish a standard analytical method for the determination of trace amounts of tin and lead in magnesium and its alloys, which could be found in neither JIS nor ISO standards, an interlaboratory testing with inductively coupled plasma atomic emission spectrometry (ICP-AES) has been put...
Saved in:
Published in | Analytical Sciences Vol. 25; no. 5; pp. 717 - 721 |
---|---|
Main Authors | , , |
Format | Journal Article |
Language | English |
Published |
Singapore
The Japan Society for Analytical Chemistry
2009
Springer Nature Singapore Japan Science and Technology Agency |
Subjects | |
Online Access | Get full text |
Cover
Loading…
Summary: | In order to establish a standard analytical method for the determination of trace amounts of tin and lead in magnesium and its alloys, which could be found in neither JIS nor ISO standards, an interlaboratory testing with inductively coupled plasma atomic emission spectrometry (ICP-AES) has been put into practice. Simple dissolution with nitric and hydrochloric acids and sample nebulization procedures were finally concluded to be suitable for standardization. The analytical methods, proposed in this study have been provided as newly established JIS standards. |
---|---|
Bibliography: | ObjectType-Article-1 SourceType-Scholarly Journals-1 ObjectType-Feature-2 content type line 23 |
ISSN: | 0910-6340 1348-2246 |